F670 HP – HIGH PRESSURE LEAK TEST INSTRUMENT FOR INDUSTRIAL QUALITY CONTROL
New electronic module, new measurement module, new interface, new accessories… All has been made to improve your quality control.
And yet, inside can be found the latest leak testing technics and technology that will guarantee you the fastest and most accurate leak testing cycle on the market: F670 HP..
F670 HP : APPLICATIONS
Automotive, domestic appliances, medical, aeronautic, defense, fluid and electrical components, food and cosmetic packaging…
F670 HP : FEATURES
- Differential pressure decay leak measurement
- Large range of leak measurement (ΔP) F.S.: 5000 Pa
- 128 programs
- Setting on line (via USB)
- 2 languages (English + other one)
- Data storage on internal memory or USB key
TEST MODES :
- ΔP (Pa)
- ΔP/t (Pa/s)
- Flow units (sccm, mm3/s, cm3/s, cm3/min, cm3/h…)
- Blockage test
- Desensitized test and others, depending on your applications
FLEXIBLE COMMUNICATION FORMAT :
- RS232: printer, modbus RTU
- Slave USB: PC supervision (WinAteq 300, sesame)
- Master USB: USB key to transfert (parameters, results, …)
- Option Fieldbus: profibus, devicenet, profinet, Ethernet/IP
ENVIRONMENT
- ROHS standard
MEASUREMENT CHARACTERISTICS :
PRESSURE DROP MEASUREMENT
| Range (Pa) | Standard accuracy* | Optional accuracy (Laboratory calibration on request) | Max. resolution (Pa) |
| 0 – 5000 | ±(2.5% Rdg + 10 Pa) | On request | 1 |
*Accuracy: Deviation + Repeatability + Hysteresis.
Rdg: Reading.
F670 HP : SPECIFICATIONS
Physical
- Weight: 11 Kg (24 lbs)
- Dimensions: (WxHxD) 537 x 157 x 380 mm
Interface
- Graphic color screen with simple and intuitive menu-driven operator controls
- SMART KEY: programmable key with user’s functions
Electrical supply
- 24V DC – 2.5A (60W)
Air supply
- Command 600 kPa (87 PSI) +/-15%
- Pressure 1 or 2 bar > Test pressure
- Clean and dry air
- Air quality standard to be applied (ISO 8573-1)
Temperature
- Temperature Operating: 5 °C to + 45 °C (41 °F to 113 °F)
- Storage: 0 °C to + 60 °C (32 °F
F670 HP : OPTIONS
- 9 inputs / 5 local outputs compatible with 5 series instruments
- 6 inputs/ 6 outputs 24 V programmable for external automation
- Mechanical Regularor 50 or 100 bar (715 or 1450 PSI)
- Electronic regulator 40 or 50 or 70 bar (580 or 725 or 1015 PSI)
- Memory card for archiving results (1 million)
F670 HP : ACCESSORIES
- Simple remote control TLC60 connection I/O
- Advanced remote control TLC600 connection master USB
- USB key for parameters access
- WinAteq 300 software
- Remote Silencer (diameter 74 mm – length 212 mm)
F670 HP : TECHNOLOGY
By far the most popular technology, differential pressure decay uses a reference volume to test your part. This helps compensate for any ambient pressure or temperature variations as they occur on both parts simultaneously.
Only a leak on the test part will result in a movement of our transducer’s membrane.
The second advantage of this method is that the accuracy does not drop with the test pressure as the transducer is measuring pressure differences between the two circuits, as opposed to the traditional pressure decay technology that measures pressure drops against the atmosphere.
Direct measurement principle
The part under test 3 and the reference part 5 are filled to an identical pressure. A differential sensor 4 measures the pressure variation between the part under test 3 and the reference part 5. In some applications, the reference part can be replaced by a cap.

Desensitized test
This mode is used for the measurement of large leaks, when the reject level required is above the full scale of the differential sensor.
The test pressure is applied to the input of the part under test 3. The measurement is performed by the pressure sensor 2.
Other type of tests are available in option

F670 HP : FACTSHEETS
Detailed information about the F670 HP including features, specifications, etc.
Contact ATEQ for sales information, general questions, press release information and marketing inquiries.